Effects of varied dimensions of surgical guides on implant angulations
Effects of varied dimensions of surgical guides on implant angulations
November 2004
By Mijin Choi DDS, Elaine Romberg PhD and Carl F. Driscoll DMD
The Journal of Prosthetic Dentistry
You can view the abstract online. A subscription is required to view the full text or it can be purchased online.
Statement of problem
Fabrication of a proper surgical guide is critical for success of implant restorations. The effects of the dimensional factors of the surgical guide on implant placement have not been studied.
Purpose
The purpose of this study was to determine the effect of varied dimensions (diameter, length, and distance between the underside of the surgical guide and the implant recipient site) of a surgical guide on the accuracy of implant angulation.
Material and methods
In this in vitro study, 240 implant recipient sites were randomly prepared using varied dimensions of a surgical guide. The varied dimensions of the surgical guide's channel and distance were: channel diameter (2, 3, 4, or 5 mm), channel length (6 or 9 mm), and distance between the underside of the surgical guide and the simulated implant recipient site (2 or 4 mm). From these varying dimensions and distances, 16 combinations of dimensions and distances were tested. For each combination, 15 simulated implant recipient site (SIRS) specimens were prepared. The deviated angulation (DA) from the midpoint of the top surface of the 1- ?Û 1-inch simulated implant recipient site (each simulated implant recipient acrylic block contained 5 SIRS of 1 ?Û 1 inch), in the right-to-left (DARL) and front-to-back (DAFB) directions, were measured in degrees using a protractor. The data was analyzed using factorial analysis of variance and Tukey's HSD test (a=.05).
Results
The DARL, in degrees, at a channel length of 9.0 mm (2.33 ¨± 1.27) was significantly smaller than at a channel length of 6.0 mm (3.0 ¨± 1.42, P=.0001). The DARL, in degrees, at a distance of 4.0 mm (2.13 ¨± 1.16) was significantly smaller than at a distance of 2.0 mm (3.16 ¨± 1.39, P=.0001). Also, a significant interaction for DARL was found between diameter and distance (P<.05). For DAFB, the varying diameters (P<.05), lengths (P=.0001), and distances (P=.0001) showed significant differences. The DAFB at a channel length of 9.0 mm (2.56 degrees ¨± 1.51) was significantly smaller than that at 6.0 mm (3.82 degrees ¨± 1.87). Significant interactions found for DAFB were: diameter by length (P=.0001), diameter by distance (F=4.547, P=.004), and length by distance (F=11.512, P=.001).
Conclusion
Within the limitations of this study, the results suggest channel length as the primary controlling factor in minimizing deviated angulations.