Abstract:
Objectives: Although bone loss around implants is reported as a complication when it progresses uncontrolled, resorption does not always lead to implant loss, but may be the result of biomechanical adaptation to stress. To verify this hypothesis, a three-dimensional finite element analysis was performed and the influence of marginal bone resorption amount and shape on stress in the bone and implant was investigated.
Material and methods: A total of nine bone models with an implant were created: a non-resorption (Base) model and eight variations, in which three different resorption depths were combined with pure vertical or conical (verticalǃÏhorizontal) resorption. Axial and buccolingual forces were applied independently to the occlusal node at the center of the superstructure.
Results: Regardless of load direction, bone stresses were higher in the pure vertical resorption (A) models than in the Base model, and increased with resorption depth. However, cortical bone stress was much lower in the conical resorption models than in both the Base and A models of the same resorption depth. An opposite tendency was observed in the cancellous bone under buccolingual load.
Under buccolingual load, highest stress in the implant increased linearly with the resorption depth for all the models and its location approached the void existing below the abutment screw.
Conclusions: The results of this analysis suggest that a certain amount of conical resorption may be the result of biomechanical adaptation of bone to stress. However, as bone resorption progresses, the increasing stresses in the cancellous bone and implant under lateral load may result in implant failure.